HMP155A-L Air Temperature and Relative Humidity Sensor

Gọi +84 904251826 để được trợ giúp

Cảm biến đo nhiệt độ và độ ẩm Campbell Scientific HMP155A-L

Overview

The HMP155A provides reliable relative humidity (RH) and temperature measurements for a wide range of applications. It uses a HUMICAP®180R capacitive thin film polymer sensor to measure RH over the 0 to 100% RH range. A PRT measures temperature over the -80° to +60°C range. This rugged, accurate temperature/RH probe is manufactured by Vaisala.

Benefits and Features

  • Well-suited for long-term, unattended applications
  • Accurate and rugged
  • Mounts to a mast, crossarm, or user-supplied pole
  • Compatible with most Campbell Scientific data loggers

Detailed Description

To reduce the current drain, power can be supplied to the HMP155A only during measurement when the sensor is connected to the data logger’s switched 12 V terminal. Data loggers that do not have a switched 12 V terminal, such as the CR510 or CR7, can use the SW12V switched 12 V device to switch power to the sensor only during measurement.

Specifications

Electromagnetic Compatibility Complies with EMC standard EN61326-1 Electromagnetic
Filter Description Sintered PTFE
Housing Body Material PC
Housing Classification IP66
Voltage Output 0 to 1 Vdc
Average Current Consumption ≤ 3 mA (analog output mode)
Operating Voltage 7 to 28 Vdc
Settling Time 2 s (at power up)
Field-Replaceable Chip or Recalibrate Recalibrate
Tip Diameter 1.2 cm (0.5 in.)
Length 27.9 cm (11 in.)
Head Height 4 cm (1.6 in.)
Body Height 2.4 cm (0.9 in.)
Body Width 2.0 cm (0.8 in.)

 

Relative Humidity

Sensing Element HUMICAP 180R
Measurement Range 0 to 100% RH (non-condensing)
Response Time
  • The response time for the RH specification is for the HUMICAP 180R at 20°C in still air with sintered PTFE filter and a 0 to 75% RH step change.
  • 20 s (63% step change)
  • 60 s (90% step change)
Factory Calibration Uncertainty
  • The factory calibration uncertainty is defined as ±2 standard deviation limits. Uncertainty is at +20°C. Small variations are possible; see also the calibration certificate.
  • ±0.6% RH 0 to 40% RH
  • ±1.0% RH 40 to 97% RH
Accuracy
  • -NOTE- Accuracy specifications include non-linearity, hysteresis, and repeatability.
  • ±1% RH (at 15° to 25°C, 0 to 90% RH)
  • ±1.7% RH (at 15° to 25°C, 90 to 100% RH)
  • ±(1.4 + 0.032 × reading) % RH (at -60° to -40°C)
  • ±(1.2 + 0.012 × reading) % RH (at -40° to -20°C)
  • ±(1.0 + 0.008 × reading) % RH (at -20° to +40°C)
  • ± (1.2 + 0.012 × reading) % RH (at 40° to 60°C)

 

Air Temperature

Sensing Element PT 100 RTD 1/3 class B IEC 751
Measurement Range -80° to +60°C
Accuracy
  • ±(0.226 - 0.0028 x temperature)°C (-80° to +20°C)
  • ±(0.055 + 0.0057 x temperature)°C (+20° to +60°C)
Entire Temperature Range Refer to graph in probe manual.